AD-4408C Weighing Indicator Page 6
3.3. General
3.3.1. Interface
Standard serial output
Output for communication with A&D peripheral devices (20
mA current loop)
CC-Link interface CC-Link ver.1.10 remote device station
3.3.2. Measuring functions
Zero adjustment
Sets the gross weight to zero by pressing the ZERO key or
a command from the CC-Link master station.
Selection of disable or enable for the operation when
unstable.
The zero value is battery backed up.
Zero adjustable range
Can be set optionally in the range of 1 to 100% of the
weighing capacity.
Zero tracking
Tracks the weight drift around the zero point to maintain
zero.
Zero tracking time
0.0 to 5.0 seconds (Can be set optionally within the range.)
Zero tracking band
0.0 to 9.9 d (Can be set optionally within the range.)
Tare
Sets the net weight to zero by pressing the TARE key or a
command from the CC-Link master station.
Selection of disable or enable for the operation when
unstable or negative.
The tare weight is battery backed up.
Tare range: Gross weight Weighing capacity
Stability detection
Turns ON the stabilization indicator when the variables of
the weight values per sampling are within the set band in
the set time.
Stability can be confirmed using the CC-Link.
Stability detection time
0.0 to 9.9 seconds (Can be set optionally within the range.)
Stability detection band
0 to 9 d (Can be set optionally within the range.)
Digital filter Cutoff frequency range: 0.07 to 11 Hz