AD-4408C Weighing Indicator Page 28
NOTE: Record the setting values of Calf17, 18 and 19 in the “Setting List” at
the end of the manual to prepare against a failure.
By changing the parameters of Calf17, 18 and 19, “zero calibration”
and “span calibration” can be adjusted optionally. (Digital span
accuracy approx. 1/5000: The accuracy varies depending on the load
cell output accuracy and the conditions of calibration.)
Except for an emergency, perform calibration with an actual load.
Calf20
0 to 1
Output when
the gross weight
is negative
Output by the manual print command when
the gross weight is negative.
0: Enables output.
1: Disables output.
0
Calf21
0 to 1
Communication
restriction
Restriction on network-related
communication such as changing station
numbers or ID numbers and outputting by
auto printing
0: Disables restriction.
1: Enables restriction.
0
Calf22
1 to 2
Header 2
Second header for serial output
1: GS / NT / TR
2: G_ / N_ / T_ (_ space)
1
CALF17
Input
voltage
CALF19
CALF18
Displayed
weight
0
0