AD-4408C Weighing Indicator Page 24
Function No.
Parameter or
Setting range
Function Description
Default
value
Calf01
0
to 3
Unit
Unit of the weight value
0: None
1: g
2: kg
3: t
4: lb (USA version)
2
Calf02
0 to 5
Decimal
point
position
Decimal point position of the weight value
0: None 123456
1: 10
1
12345.6
2: 10
2
1234.56
3: 10
3
123.456
4: 10
4
12.3456
5: 10
5
1.23456
0
Calf03
1 to 6
Minimum
division
Minimum division (d) of the weight value
1: 1
2: 2
3: 5
4: 10
5: 20
6: 50
1
Calf04
1 to 999999
Weighing
capacity
Weighing capacity of the weighing
instrument
Weighing is possible up to the value of this
setting plus 8 d. If the value exceeds this,
overflow will occur and will not be
displayed.
The decimal point position is the same as
the setting of Calf02.
20000
Calf05
0 to 100
Zero
adjustment
range
Range to enable zero adjustment by the
ZERO key or a command from the CC-Link
master station.
Expressed in percent of the weighing
capacity with the calibration zero point as
the center, when performing power-ON
zero, with the initial zero point as the
center.
For example, if this is set to 2, the value in
the range of ±2% of the weighing capacity
with the calibration zero point at the center
will set to zero.
2