AD-4408C Weighing Indicator Page 36
9.1.2. Output data
The data format used is A&D standard format. This is a format to be used with an A&D
printer or remote display and consists of two headers, data, unit and a terminator.
A&D standard format
ASCII code Hexadecimal Description
ST [53 54] Stable
US [55 53] Unstable
Header 1
OL [4F 4C] Overload
GS [47 53] Gross weight
NT [4E 54] Net weight
Header 2
TR [54 52] Tare weight
Separator , [2C]
0 to 9 [30 to 39]
+ [2B]
- [2D]
SP (space) [20]
Data (ASCII code)
. [2E]
SP SP [20 20] No unit
SP g [20 67] g
kg [6B 67] kg
SP t [20 74] t
Unit (five types)
lb [6C 62] lb (USA version)
A&D standard format examples
Gross weight
Net weight
Tare weight
Data with a decimal point
Positive overflow
Negative overflow
Unstable data
“Output OFF” data
NOTE: The decimal point position is the same even when overflow occurs.