Page 17 AD-4408C Weighing Indicator
6.1.3. Tare
Tare is a function to store the gross weight as the tare value and set the net weight to
zero.
The tare weight is stored in the battery backup RAM and is maintained, even if the
power is disconnected.
Clearing the tare weight is performed using the F key assigned to clear the tare weight.
Alternately, apply power to the indicator while holding down the TARE key, or with the indicator in
the OFF mode, press and hold the TARE key and press the ON/OFF key, to clear the tare weight.
Functions related to tare
z Calf10: Enables or disables tare when unstable.
z Calf11: Enables or disables tare when the gross weight is negative.
6.1.4. F key
Assign a function to the F key in the general functions.
Functions related to the F key
zfncf02: Assigns a function to the F key from the functions below:
0: None
1: Manual print command
2: Hold
3: Alternate switch (Read the description below.)
4: Momentary switch (Read the description below.)
5: Clear the tare weight
6: Clear the zero value
z
Calf15: Enables or disables clearing the zero value.
zCalf20: Enables or disables the print command when the gross weight is negative.
Alternate switch and momentary switch
zBy assigning these switches to the F key, the ON/OFF status of the F key can be
transmitted to the CC-Link master station. This is useful when building a CC-Link
network or performing maintenance.
zIn the case that the station number is 1 and the number of occupied stations is 4, the
ON/OFF status is transmitted to RX006F of the address map (In sync with
display) . (ON: 1, OFF: 0)
zThese switches perform as below:
Alternate switch
Press the switch once and release the switch to turn ON or OFF.
Press the switch again to turn OFF or ON.
Momentary switch
Only when the switch is being pressed, the F key is ON. When released, OFF.