AD-4408C Weighing Indicator Page 52
10.2.4. Checking the CC-Link
With Chc CC displayed, press the ENTER key to enter the CC-Link check mode.
Press the ∧ or ∨ key to change the address.
Press the CAL key to transmit the request flag of initialization.
Address Remote I/O Buffer memory
05 to 0C RY0000 to RY007F 160 to 167
0D to 1C RWw0000 to RWw000F 1E0 to 1EF
25 to 2C RX0000 to RX007F 0E0 to 0E7
2D to 3C RWr0000 to RWr000F 2E0 to 2EF
10.2.5. Checking the testing terminal
The Chc r5 display is for checking the testing terminal. Do not use it.
10.2.6. Checking the A/D (load cell)
With Chc ad displayed, press the ENTER key to enter the A/D check mode. The load cell
output value is displayed in mV/V.
The example below is when the internal count is 1.23456 mV/V.
If the internal count exceeds ±7mV/V, the load cell may be damaged or the load cell may
be connected incorrectly. Refer to “10.4. Checking the Load Cell Connection Using the
Digital Multimeter” to check the connection.
10.2.7. Checking the internal count
With Chc in displayed, press the ENTER key to enter the internal count check mode.
The internal count value (display value times 10) is displayed.
The example below is when the internal count is 123.
1 . 23456
123.
88. 8888
Remote READY flag
Error status flag
Reply flag of initial data setting
Request flag of initialization
CPU normal operation
Link status
2 digits: Address (00 to 3F)
4 digits: Data (0000 to FFFF)