Page 33 AD-4408C Weighing Indicator
8.2. Basic Functions (fncf)
Function No.
Parameter or
Setting range
Function Description
Default
value
Ffncf01
000000 to
111111
Disable
key switch
Each digit of the setting corresponds to a key
switch. Only available in the weighing mode.
0: Does not disable the key switch function.
1: Disables the key switch function.
Key assignment
6
th
5
th
4
th
3
rd
2
nd
1
st
000000
(binary)
fncf02
0 to 6
F key
Assigns a function to the F key.
0: None
1: Manual print command
2: Hold
3: Alternate switch
4: Momentary switch
5: Clear the tare weight
6: Clear the zero value
NOTE: Clearing the zero value can be
enabled or disabled by Calf15
0
fncf03
1 to 3
Display
update
rate
1: 20 times/second
2: 10 times/second
3: 5 times/second
1
fncf04
0 to 3
display
Assigns a function to the display.
0: None
1: Zero tracking in progress
2: Alarm (Zero range setting error, over,
low battery)
3: F key status
0
fncf05
0 to 9
Digital
filter
Selects a cutoff frequency.
0: None
1: 11.0 Hz
2: 8.0 Hz
3: 5.6 Hz
4: 4.0 Hz
5: 2.8 Hz
6: 2.0 Hz
7: 1.4 Hz
8: 1.0 Hz
9: 0.7 Hz
8
fncf06
1 to 10
Frequency
divider
ratio
Divides digital filter cutoff frequency by the
setting.
1
fncf07
1 to 2
Hold
1: Normal hold
2: Peak hold
1
ZERO TARE
Net/B/G
NET GROSS
ON/OFF
ESC
ENTER
Comentarios a estos manuales