AD-4408C Weighing Indicator Page 18
6.1.5. display
Assign a function to the display in the general functions.
Functions related to the display
zfncf04: Assigns a function to the display from the functions below:
0: None
1: Zero tracking in progress
2: Alarm (Zero range setting error, over, low battery)
3: F key status (Read the description below.)
F key status
zEnables when alternate switch or momentary switch is selected for fncf02.
The display turns ON when the F key is ON and turns OFF when the F key is OFF.
6.1.6. Memory backup
The indicator has two kinds of memory.
EEPROM Memory used to store important data, without power
supplied, that the occurrence of re-writing is seldom.
Data example: Calibration data, Function data
Battery backup RAM Memory used to store temporary data that the occurrence
of re-writing is often.
Data example: Zero value, Tare weight
Memory backup
Data Method Description
Calibration data
Function data
EEPROM Maintained regardless of battery condition.
Zero value
Tare weight
Battery
backup RAM
The life of the battery is approximately 10 years
at 25°C with the power disconnected.