Page 55 AD-4408C Weighing Indicator
10.3.2. All data initialization
Step 1 Remove the cover concealing the CAL key and press the CAL key.
Cal is displayed to indicate that the indicator enters the calibration mode.
Step 2 Press the ENTER key to display Cal5et . The indicator enters the mode of
calibration with an actual load. To return to the weighing mode, press the ESC
key.
Step 3 Press the ∧ or ∨ key to select the all data initialization mode ( iniallLL ) and
press the ENTER key.
Step 4 All the status LEDs blink to prompt a confirmation. To initialize, hold down the
ENTER key for three seconds or more.
After initialization, the indicator is reset and all the display segments are ON.
And the indicator enters the weighing mode.
To exit from this mode without performing initialization, press the ESC key.