AD-4408C Weighing Indicator Page 60
11.4. Calibration-Related Functions
Function No.
Parameter or
Setting range
Description
Default
value
User
setting
Calf01
0
to 3
Unit of the weight value
0: None
1: g
2: kg
3: t
4: lb (USA version)
2
Calf02
0 to 5
Decimal point position of the weight value: 10
n
0
Calf03
1 to 6
Minimum division (d) of the weight value
1: 1
2: 2
3: 5
4: 10
5: 20
6: 50
1
Calf04
1 to 999999
Weighing capacity of the weighing instrument
The decimal point position is the same as
the setting of Calf02.
20000
Calf05
0 to 100
Range to enable zero adjustment by the
ZERO key or a command from the CC-Link
master station.
Expressed in percent of the weighing
capacity with the calibration zero point as
the center
2
Calf06
0.0 to 5.0
Zero tracking time (Unit: second)
When 0.0, zero tracking will not be performed.
0.0
Calf07
0.0 to 9.9
Zero tracking band (Unit: 0.1 d)
When 0.0, zero tracking will not be performed.
0.0
Calf08
0.0 to 9.9
Stability detection time (Unit: second)
When 0.0, stability detection will not be
performed.
1.0
Calf09
0 to 9
Stability detection band (Unit: d)
When 0, stability detection will not be
performed.
2
Calf10
0 to 1
Tare and zero adjustment when the weight
value is unstable.
0: Disables both functions.
1: Enables both functions.
1
Calf11
0 to 1
Tare when the gross weight is negative.
0: Disables tare.
1: Enables tare.
1
Calf12
0 to 1
Standard serial output if the weight value
overflows and is unstable.
0: Disables output.
1: Enables output.
1