Page 27 AD-4408C Weighing Indicator
Function No.
Parameter or
Setting range
Function Description
Default
value
Calf13
1 to 3
Over at
negative
gross weight
Condition to judge over when the gross
weight is negative.
A/D negative over or
1: Gross weight < -999999
2: Gross weight < negative weighing capacity
3: Gross weight < -19 d
1
Calf14
1 to 2
Over at
negative net
weight
Condition to judge over when the net
weight is negative.
Gross weight negative over or
1: Net weight < -999999
2: Net weight < negative weighing capacity
1
Calf15
0 to 1
Clear the zero
value
Select whether or not to clear the zero
value.
0: Disables.
1: Enables.
1
Calf16
0 to 1
Zero upon
power-ON
Select whether or not to perform zero
upon power-ON.
0: Disables.
1: Enables.
0
Calf17
-7.00000 to
7.00000
Input voltage
at zero
Input voltage from a load cell at zero.
Unit: mV/V
In zero calibration during the calibration
with an actual load, this value is
determined.
0.00000
Calf18
0.00001 to
9.99999
Input voltage
at span
Input voltage from a load cell at span
Unit: mV/V
In span calibration during the calibration
with an actual load, this value and the
value of Calf19 are determined.
2.00000
Calf19
1 to 999999
Calibration
weight value
corresponding
to Input
voltage at
span
The calibration weight value corresponding
to the Input voltage at span of Calf18.
When performing digital span, Calf17,
18 and 19 are required.
The decimal point position is the same as
the setting of Calf02.
20000