- 42 - AD-4430B
When the module does not operate properly, write the required items in the table below and contact
your local A&D dealer.
Item
User’s Usage Circumstances
Model Number, Rated, Measurement
Value etc.
Description
Connection method
4-wire connection
6-wire connection
When using the 4-wire connection,
connect between EXC+ and SEN+ and
between EXC- and SIG-.
Model number of load cells
Rated capacity of load cells [Unit ]
Rated output of load cells [mV/V]
Allowable overload of load
cells
[%]
Number of load cells used [pieces]
Use of summing box
Length of the extension cable [m]
Length from the module to the summing
box.
Initial load on the module [Unit ]
Weighing capacity of the
module
[Unit ]
Write all digits including decimal figures.
Example : 0.002 kg
Weighing capacity of the
module
[Unit ]
Write all digits including decimal figures.
Example : 10.000 kg
Load cell output during initial
load (when no load is applied)
[mV/V]
From -0.1 mV/V to rated sensitivity value
(With the first load)
Relation between load cell
output and load of mass of
capacity (or chosen mass )
Load :
[Unit ]
Load cell output :
[mV/V]
When loaded to capacity, the output
value of the initial load + the rated output
value of the load cell. (It must be within
allowable overload.)
Measurement point Measurement contents Measurement result
EXC+ SEN+ A drop voltage of the load cell cable on EXC+ side. [mV]
EXC+ EXC- Input voltage of load cell [V]
SEN- EXC- A drop voltage of the load cell cable on EXC- side. [mV]
SIG- EXC- Center point voltage [V]
SIG+ SIG- Output voltage of load cell [mV]