A&D AD-4430B Especificaciones Pagina 34

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 33
- 32 - AD-4430B
5.4.3. The List Of The Basic Functions
Function No.
Parameter
and range
Function name Description
Default
value
Fnc01
0000 to 1111
Disable key
switch
Each digit of the setting corresponds to a key switch.
Only available in the weighing mode.
0 : Enables Key assignment to each binary digit.
1 : Disables 4th 3rd 2nd 1st
0000
(2binary)
Fnc02
0 to 7
key
0 : None
1 : Manual print command
2 : Hold
3 : Alternate switch
4 : Momentary switch
5 : Changing between gross weight and net weight
6 : Clear the tare weight
7 : Clear the zero value
NOTE: Clearing the zero value can be enabled or
disabled by CAL15.
0
Fnc03
1 to 3
Display update
rate
1 : 20 times/second
2 : 10 times/second
3 : 5 times/second
1
Fnc04
0 to 9
x display
0 : None
1 : Zero tracking in progress
2 : Alarm (Zero range setting error, over)
3 : key status
4: : Zero band
5 : HI output (Over the upper limit value)
6 : OK output (Between upper and lower limit values)
7 : LO output (Below the lower limit value)
8 : User input 1
9 : User output 1
0
Fnc05
0 to 16
Digital filter
Selects a cutoff frequency.
0: None
1: 100.0 Hz 9: 7.0 Hz
2: 70.0 Hz 10: 5.6 Hz
3: 56.0 Hz 11: 4.0 Hz
4: 40.0 Hz 12: 2.8 Hz
5: 28.0 Hz 13: 2.0 Hz
6: 20.0 Hz 14: 1.4 Hz
7: 14.0 Hz 15: 1.0 Hz
8: 10.0 Hz 16: 0.7 Hz
15
Fnc07
1 to 3
Hold
1 : Normal hold
2 : Peak hold
3 : Averaging hold
1
Fnc08
-99999 to
99999
Set value of
near-zero
The reference value for near-zero.
Decimal point position is linked to CAL02.
10
Vista de pagina 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 55 56

Comentarios a estos manuales

Sin comentarios