- 16 - AD-4430B
5.1.9. The Hold Function
There are three types of hold functions which can be used for different purposes.
Normal hold
Holds the value displayed at the time the hold command was received.
Peak hold
Holds the maximum value reached after the hold command was received.
The value will be refreshed if it increases again.
Averaging hold
This function averages weighing data over a certain period of time and then holds the result.
It is useful for measuring things that are difficult to weigh such as an animal that won’t settle
down, or for averaging out the weight of an object in an unstable state. In addition, it can
reduce the effects of breezes which the digital filter cannot eliminate.
Functions Related To The Hold Function
Fnc07(Hold) : The type of hold function can be selected.
To set operating conditions for the hold function : Set the averaging time length, standby
time, or start and stop conditions with HLd01 to 07. (Will not affect normal holds)
Peak hold
Averaging hold
Illustration 6 Peak hold / Averaging hold
Waiting time (Busy state)
(Start standby timer)
veraging (Busy state)
(Averaging time)
Hold
Hold start command
Automatic start condition
veraging star
veraging finish Hold cancellation
command
Automatic cancellation condition
Standard
state
Standard
state
Waiting time (Busy state)
(Start standby timer)
Hold
Hold start command
Automatic start condition
Hold cancellatio
command
Automatic cancellation condition
Standard
state
Standard
state
Refresh
(Busy state)
Hold
Judgment star