A&D AD-4430B Especificaciones Pagina 30

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 29
- 28 - AD-4430B
5.3.7. The Function Related To The Linearization Function
Confirm and change linearity settings.
To use this function, select in the same way as calibration-related functions are
selected.
Function No.
Parameter
and range
Function name Description
Default
value
Lnr01
1 to 4
Number of input
points
Number of points where linear input was done.
The linear-zero input is included as one point.
Digital linearization is not performed when the set
value is between 1 and 2.
0
Lnr02
-7.0000
to 7.0000
Linear-zero Voltage for linear-zero input. Unit : mV/V. 0.0000
Lnr03
0 to 99999
Linear 1
Mass value
The mass value for linear 1 input.
The decimal point position depends on CAL02.
0
Lnr04
0.0000
to 9.9999
Linear 1
Span
The span voltage between linear-zero and linear 1
input. Unit : mV/V.
0.0000
Lnr05
0 to 99999
Linear 2
Mass value
The mass value for linear 2 input.
The decimal point position depends on CAL02.
0
Lnr06
0.0000
to 9.9999
Linear 2
Span
The span voltage between linear-zero and linear 2
input. Unit : mV/V.
0.0000
Lnr07
0 to 99999
Linear 3
Mass value
The mass value for linear 3 input.
The decimal point position depends on CAL02.
0
Lnr08
0.0000
to 9.9999
Linear 3
Span
The span voltage between linear-zero and linear 3
input. Unit : mV/V.
0.0000
Lnr09
0 to 99999
Linear 4
Mass value
The mass value for linear 4 input.
The decimal point position depends on CAL02.
0
Lnr10
0.0000 to
9.9999
Linear 4
Span
The span voltage between linear-zero and linear 4
input. Unit : mV/V.
0.0000
Vista de pagina 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 55 56

Comentarios a estos manuales

Sin comentarios