- 38 - AD-4430B
7.3. Check Mode
The check mode checks the performance of the display, key switches and external I/O.
7.3.1. Entering The Check Mode
Step 1
Press the key while pressing and holding the key ( + ) to
display .
If returning to the weighing mode, press the key.
Step 2
Press the
key while pressing and holding the key (
+ ) to
display of the check mode. Press the key to display the check item
Step 3
Select the check item using the
key. Press the key to enter to it.
Press the key when exiting it
Display Item To Be Checked
Key check
BCD check
A/D converter output check (Load cell)
Internal count
Program version
Serial number
Program checksum
Memory checksum (EEPROM)
CALF check (CAL01 to CAL28)
7.3.2. Verifying The Switch Operation
When pressing the key, the corresponding segment moves.
If stopping the current check mode, press the key twice.
7.3.3. Monitoring The A/D Converter ( For Load cell Output)
The load cell output voltage rate is displayed in unit of mV/V .
ex. If internal count is 1.2345 mV/V ,
If value is above ±7 mV/V, a damage and connection error of the load cell may cause . Refer
to "7.5. Verifying The Load cell Connections Using Multimeter".