5.3.6. The Function Related To The Calibration
Step 1 Press the + key to enter to the calibration mode and display .
Press the key to start the calibration and display .
Select using the key and press the key.
Step 2 Select using the key and press the key.
Step 3 Select a desired function number using the key and press the key.
The current data is displayed.
Step 4 When changing data, two methods of parameter selection and digital input depending on
the function are available.
Type Description of method to change data
Parameter selection
Only the available parameter is displayed and blinks.
Select a number using the key.
Digital input
All the digits are displayed and a digit to be changed blinks.
Select a digit using the
key and
change the value using the key.
After changing data, press the key. The next function number is displayed.
When the value is not to be changed, press the key to return to the function
number display.
Step 5 Press the key to store new data in FRAM and is displayed.
Press again the key to return to the weighing mode.
The blinking decimal point means that the current value is not the weight value.
If a data exceeding the available range is inputted, is displayed and data is
canceled.