A&D GR Series Instruction Manual
GB 70
Display Error code Description of the error
Weighing pan Error
This is a warning that the weight value is too light.
Confirm that the weighing pan and the pan support are
properly installed.
Unit weight, 100% weight error
The unit weight of the sample is very light in the counting
mode, or the 100% sample is too light in percent mode.
The balance can not calculate it.
Increase the unit weight or 100% weight.
ARA Zero error
The ARA (Automatic Response Adjustment) can not be
performed, because there is something on the pan.
Remove all matter from the pan. Press the [CAL] key to
return to the weighing mode.
ARA Unstable error
The ARA (Automatic Response Adjustment) can not be
performed because of unstable weighing value.
Correct the environment for the balance. Press the [CAL]
key to return to the weighing mode.
Full memory
The memory data has reached 200 items.
When data is deleted, new data can be stored.
Memory data error
The memory data is lost.
Clear all memory data.
Unit weight information
This is advice regarding the sample number that is needed
to set the unit weight. When the unit weight is computed
and the sample number is too few, the required number is
displayed for counting accuracy.
Count and place the samples on the pan. Press the
[PRINT] key to store the correct value.
Other errors
If you cannot cancel the error yourself, request service from your supplier of the balance or option,
alternatively, the A&D service group can help.