10. ID Number and GLP Report
10.1. Setting of the ID Number
• The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.
• The ID number is output on the “Calibration Report”, “Calibration Test Report” and “Title Block”.
• The GLP output format is selected at the “GLP output (inFo)” of the “Function Table”.
• The balance can output the following reports for GLP:
“Calibration Report” using the internal weight.
“Calibration Report” using an external weight.
“Calibration Test Report” using the internal weight.
“Calibration Test Report” using an external weight.
“Title block” and “End block” for weighing data.
1. Press and hold the [RANGE] key to display bASFnc.
2. Press the [RANGE] several times to display id.
3. Press the [PRINT] key. You can set the ID number using the following keys:
[RANGE] key Increments the digit.
[RE-ZERO] key The key to select the character of the digit. Refer to the following table for the
“Display Character Set”.
[PRINT] key The key to store a new ID number and proceed to the next class of the function
table.
[CAL] key The key to cancel the new ID number and proceed to the next class of the
function table.
4. Press the [CAL] key to return to the weighing mode.
Display Character Set