GB 32
This function returns the following parameters to factory settings.
• Calibration data.
• Function table.
• The 100% weight.
• The data that is stored in the balance using data memory function.
• External calibration weight and target weight.
• Switch settings for “Permission or Prohibition”.
Operation
1. Turn off the display.
2. Press the [ON:OFF] key while the [RANGE] key and
[PRINT] key are pressed and held. Then the balance displays PS.
3. Press the [RANGE] key to display CLr.
4. Press the [PRINT] key. (If you want to cancel this operation,
press the [CAL] key.)
5. Press the [RE-ZERO] key.
6. Press the [PRINT] key to initialize the balance.
The balance will automatically return to weighing mode.
+
Press and hold
these keys
Press the key
7.2. Initialising the Balance