A&D GR Series Instruction Manual
GB 16
Note: This function automatically calibrates the balance, when the balance detects an ambient
temperature change.
When the balance detects a change in ambient temperature, this indicator
blinks and automatic self calibration is required. If the balance is not used for
several minutes with this indicator blinking, the balance performs automatic self
calibration. The environment will affect the time that the indicator blinks.
This display means “the balance is measuring calibration data”. Do not allow
vibration or draughtrs to affect the balance while this indicator is displayed.
You can use the balance while the indicator blinks. But, it is recommended that to maintain the best
accuracy, stop using the balance and confirm that there is nothing on the pan when the indicator starts
blinking. Allow the balance to self-calibrate or initiate the calibration by pressing the [CAL] button.
Note: Do not place anything on the weighing pan during automatic calibration.
Control of Automatic Self Calibration
Automatic self calibration is controlled by a parameter in “Permission or Prohibition”. Refer to
Section “Function Switch and Initialization”.
Automatic Self Calibration
0 : Not used
1 : Used
3.2. Automatic Self Calibration