10.6. Calibration test report using an external weight
English
GB 47
Key Operation
1. Press and hold the [CAL] key until displaying CC out. Release the key.
2. The balance displays CC 0.
• If you want to change the calibration weight value, proceed to step 3.
• If you use the stored calibration weight value in the balance, proceed to step 4.
3. Press the [RANGE] key and adjust target weight using the following keys.
[RE-ZERO] key The key to set the value of the digit selected.
[RANGE] key The key to select the digit to change value.
[PRINT] key The key to store a new weight value and return to step 2.
[CAL] key The key to cancel this change and return to step 2.
4. Press the [PRINT] key. The zero point is measured and this value is displayed.
5. Place the calibration weight on the pan and press the [PRINT] key. The weight is measured and
this value is displayed.
6. Remove the weight after End is displayed.
7. If the calibration report is output, GLP is displayed and the GLP data is output.
8. The balance returns to the normal weighing mode automatically.