3.3. Calibration Using the Internal Weight
English
GB 17
Operation
1. Connect the AC adapter and warm up the balance for at least
one hour with nothing on the weighing pan.
2. Press the [CAL] key to start calibration.
3. The balance displays CAL in and performs calibration.
Prevent vibration and draughts from affecting the balance.
4. If the “GLP output (inFo)” of the “Function Table” is set to “/
or “2”, “Calibration Report” is output from RS-232C interface.
5. The balance will automatically return to the weighing mode
after calibration.
6. Test the accuracy of weighing using the calibration test function
or by using a certified test weight.
Control of this Calibration
Calibration using the internal weight is controlled by a parameter
in “Permission or Prohibition”. Refer to Section “Function Switch
and Initialization”.
Calibration Using the Internal Mass
0 : Not used
1 : Used