A&D SV-10A Manual de instrucciones Pagina 54

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 102
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 53
52
9-2 Details of the Function Items (For all models)
Function item Parameter Description
5vtyp Htype
Portable typeThis function can use: H-fnc, 5t-b, Hld-t, Cp, bep, Cp Hi, Cp lo
With the SV-H series, it is shipped with this setting.
Change type
atype
Stand typeThis function can not use: H-fnc, 5t-b, Hld-t, Cp bep, Cp Hi,
Cp lo With the SV-A series, it is shipped with this setting.
Cond 0 Follows the viscosity changes quickly. (Prone to vibration)
Condition 1
2 Follows the viscosity changes slowly. (Stable values)
Unit 0
M
mPas
Unit upon power-on 1 Pas
2 CcP
3 P
°C
4
m
mPas
5 Pas
6 ccP
7
Viscosity
P
Temper-
ature
°F
Factory setting:
SV-1A/1H/10A/10H=0
SV-100A/100H=1
pnt 0 Dot
Decimal point 1 Comma
With "Comma" selected, the
separator for CSV format will
be ";" (semicolon).
fnc 0 Switches viscosity units.
1
Switches between the temperature display and the measurement
elapsed time display.
MODE key function
during
measurement
2 Do not display the digits that are beyond the decimal point.
prt 0 Key mode
Press the PRINT key to output data.
Data output mode 1 Auto print mode
Outputs automatically when
the STOP key ends the
measurement.
2 Stream mode
Continuous output during
measurement. Outputs the viscosity
only when D.P. format is selected.
type 0 A&D standard format For AD-8121B MODES 1&2
Data output format 1 D.P. format For AD-8121B MODE 3
2 CSV format For a personal computer
3 RsVisco format For graphing program RsVisco
5-at 0 No output
Measurement
elapsed time output
1 Output
Available only for D.P.
format
5-td 0 No output
Date/time output 1 Output
Available only for D.P. and
CSV formats
5-ed 0 No output
Other output 1 Outputs remarks.
2
Outputs remarks, Device ID
information and signature.
Available only for D.P. format
3 Outputs ID number. Available only for CSV format
pU5e 0 No pause
Pause at data output 1 Pause (Approx. 2 seconds)
erfnc 0 Usually use this parameter.
Reserved
7
H-fnc 0 Off
1 Auto hold function Function of
portable type
2 Timer mode function
Factory setting