8-2 Calibration Procedure (For all models)
Note
As to the correction value used for one-point calibration and two-point calibration, enter the
product of the absolute viscosity value and the density of the standard viscosity fluid. For
details, refer to "5-5-2 At Calibration".
After calibration, check the values, comparing the product described above with the
displayed value.
Kinetic viscosity = From this, Viscosity × Density =
Example 1: To calibrate the viscometer using a standard viscosity fluid:
Using the calculation sheet, calculate the value used for calibration.
(1) Check the kinetic viscosity and the viscosity at the temperature when the calibration is
performed.
Here, 1011 mm
2
/s for the kinetic viscosity and 889 mPas for the viscosity at 20°C
as an example.
(2) Substitute the values above to obtain the value for .
781
781 mPas is obtained as a correction value used for calibration.
Example 2: To calibrate using a standard viscosity fluid with known values of viscosity and density.
In this example, a standard viscosity fluid with a viscosity of 889 mPas at 20°C is
used.
(1) Check the viscosity value and the density of the standard viscosity fluid at the
temperature when the calibration is performed..
Here, 889 mPas for the viscosity and 0.878 for the density at 20°C as an example.
(2) Substitute the values above to obtain the value for Viscosity × Density.
889 × 0.878 781
781 mPas is obtained as a correction value used for calibration.
889
2
1011
Viscosity
Density
Viscosity
2
Kinetic viscosity is obtained.
Viscosity
2
Kinetic viscosity