❚ Fast and uniform heating with halogen lamp and innovative SRA
(Secondary Radiation Assist) technology
❚ High repeatability with SHS (Super Hybrid Sensor) technology
❚ WinCT-Moisture for suggestion of optimum heating temperature
(RsTemp)
and real-time graphical display (RsFig) (standard for MS & MX)
❚ Sodium Tartrate Dihydrate for accuracy checking
❚ Calibration function for heating temperature with AX-MX-43
(MS/MX only)
❚ Memory function
❚
Five measurement programs
(Standard, Quick, Automatic, Timer and Manual Mode)
❚ Four heating mode for MS/MX/MF
(Standard, Quick, Step and Ramp Heating Mode)
❚ Large VFD display for clear and easy to see
❚ Easy handling of the unit
❚ Low maintenance cost assured
❚ Progress window for heating check
❚ Self check function
❚ Quick reference card
❚ Standard RS232C
❚ Conformity to GLP / GMP / GCP / ISO with date/time
❚ ID, calibration data and check record output
Model
Measurement Method
Max Sample Weight Capacity
Weight Resolution
Moisture Content Display
Moisture Content over 1g
Accuracy over 5g
Heating Technology
Drying Temperature
Measurement Programs
Measurement Mode
Display Type
Sample Pan Size
MS-70
71g
0.0001g
0.001%/0.01%/0.1%
0.05%
0.01%
30 — 200C
MX-50
0.001g
0.01%/0.1%
0.10%
0.02%
MF-50
51g
0.002g
0.05%/0.1%/1%
0.20%
0.05%
50 — 200C
ML-50
0.005g
0.1%/1%
0.5%
0.1%
400W straight halogen lamp heating system
with SRA filter and SHS weighing technology
Halogen lamp (Straight type, 400 Watt max, 5000 hours)
Standard Mode / Automatic Complete Mode /
Timer Mode / Manual Mode / Quick Mode
Wet base / Dry base / Solid content / Ratio / Weight
Large VFD
Ø 85mm
WinCT
-
Moisture
WinCT
-
Moisture
WinCT–Moisture consists of RsTemp software to determine the heating temperature
and RsFig software for graphics.
Measurement example
Soybean flour
RsTemp RsFig
Sodium tartrate dihydrate
Moisture Analyzers