z Compliance with ISO8655 or any other specifications based on the
gravimetric method”
z Select from three models covering a wide volumetric range
z Easy test and data management using special WinCT-Pipette software
z Standard liquid thermometer and evaporation trap to ensure as precise
measurements as possible
z Includes a calibration weight and tweezers for the balance
Pipette Accuracy
Testers
*
4
Pipette Nominal
Volume
ISO8655 Requirements
(Gravimetric Method)
Maximum Permissible Error
Balance Minimum
Weighing Value
Accuracy
(Systematic Error)
Repeatability
(Random Error)
±
1.0
1.0
0.8
0.8
0.8
0.8
0.8
0.8
0.6
±µL
0.2
0.5
0.8
1.6
4.0
8.0
16
40
60
%
0.5
0.4
0.3
0.3
0.3
0.3
0.3
0.3
0.3
µL
0.1
0.2
0.3
0.6
1.5
3.0
6.0
15.0
30.0
AD
-
4212B
-
PT
AD-4212A-PT
mg
0.01
0.1
(µL)
20
50
100
200
500
1000
2000
5000
10000
Daily inspection, simplified verification 1
FX
-
300i
-
PT
Corresponding Models
Pipette Specifications in accordance with ISO8655
*
6
*
5
*
4 The maximum volume selectable for variable volume pipettes
*
5 The AD-4212B-PT can be used for the pipette volume range from 20 µL to 10000 µL
*
6 The minimum weighing value, 1mg, corresponds to approximately 1 µL. If a pipette volume is
1000 µL, a test can be performed with a resolution of 0.1%. If 200 µL, 0.5%.
Note) Make sure that the measurement environment is free from vibration, drafts and air from air conditioners.