AD-4402 Page 91
RS-422 Connections
Settings RS-422 [03 f-11] [1]
Address Number 0 [03 f- 8] [0].
RS-485 4 Wire Connections
Settings RS-422 [03 f-11] [1]
Address Number [03 f- 8] [without 0].
Host computer
RDA 3
RDB 4
SG 5
FG 6
SDA 1
SDB 2
A
D-4402
TERMINATOR
ON
OFF
Terminator resistor
100 ~ 120 1/2W
Turn on terminator
switch
The host computer may include
the terminator resistor.
The host computer may have
inverse polarity terminals.
Use twisted-pair wire.
Use shield cable, if it is needed.
RDA 3
RDB 4
SG 5
FG 6
SDA 1
SDB 2
A
D-4402
No.1
Host computer
Terminator resistor
100 ~ 120 1/2W
The host computer may
include the terminator
resistor.
The host computer may have
inverse polarity terminals.
Use twisted-pair wire.
Use shield cable, if it is needed.
TERMINATOR
ON
OFF
A
D-4402
No.32
TERMINATOR
ON
OFF
Connect a terminator
resistor at the indicator that
is the longest distance from
the host computer.
RD
A
RDB
SG
FG
SDA
SDB
RD
A
RDB
SG
FG
SDA
SDB
RD
A
RDB
SG
FG
SDA
SDB