A&D Series 57ZZ Manual de usuario Pagina 27

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 44
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 26
Acquisition Controls for Digitizing
Oscilloscopes
Digitizing oscilloscopes have settings that let you
control how the acquisition system processes a
signal. Look over the acquisition options on your
digitizing oscilloscope while you read this descrip-
tion. Figure 37 shows an example of an acquisition
menu.
Acquisition Modes
Acquisition modes control how waveform points are
produced from sample points. Recall from the first
section that sample points are the digital values that
come directly out of the Analog-to-Digital-Converter
(ADC). The time between sample points is called the
sample interval. Waveform points are the digital
values that are stored in memory and displayed to
form the waveform. The time value difference
between waveform points is called the waveform
interval. The sample interval and the waveform
interval may be, but need not be, the same. This fact
leads to the existence of several different acquisition
modes in which one waveform point is made up
from several sequentially acquired sample points.
Additionally, waveform points can be created from a
composite of sample points taken from multiple
acquisitions, which leads to another set of acquisi-
tion modes. A description of the most commonly
used acquisition modes follows.
Sample Mode: This is the simplest acquisition
mode. The oscilloscope creates a waveform point
by saving one sample point during each wave-
form interval.
Peak Detect Mode: The oscilloscope saves the
minimum and maximum value sample points
taken during two waveform intervals and uses
these samples as the two corresponding wave-
form points. Digitizing oscilloscopes with peak
detect mode run the ADC at a fast sample rate,
even at very slow time base settings (long wave-
form interval), and are able to capture fast signal
changes that would occur between the waveform
points if in sample mode. Peak detect mode is
particularly useful for seeing narrow pulses
spaced far apart in time.
Hi Res Mode: Like peak detect, hi res mode is a
way of getting more information in cases when
the ADC can sample faster than the time base
setting requires. In this case, multiple samples
taken within one waveform interval are averaged
together to produce one waveform point. The
result is a decrease in noise and an improvement
in resolution for low-speed signals.
Envelope Mode: Envelope mode is similar to peak
detect mode. However, in envelope mode, the
minimum and maximum waveform points from
multiple acquisitions are combined to form a
waveform that shows min/max changes over
time. Peak detect mode is usually used to acquire
the records that are combined to form the enve-
lope waveform.
Average Mode: In average mode, the oscilloscope
saves one sample point during each waveform
interval as in sample mode. However, waveform
points from consecutive acquisitions are then
averaged together to produce the final displayed
waveform. Average mode reduces noise without
loss of bandwidth but requires a repeating signal.
A special note about DPO acquisition: The Digital
Phosphor Oscilloscope has a high display sample
density and an innate ability to capture intensity
(Z-axis) information. With its intensity axis, the DPO
is able to provide the same type of 3-dimensional,
real-time display that analog scopes are known for.
As you look at the waveform trace on a DPO, you can
see brightened areas. These are the areas where the
signal occurs most often. This makes it easy to
distinguish, for example, the basic signal shape from
a transient that occurs only once in a while. The
basic signal would appear much brighter.
DPOs also include all the acquisition modes
described above.
21
Figure 37. Example of an acquisition menu.
Vista de pagina 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 43 44

Comentarios a estos manuales

Sin comentarios