A&D OP-05 Manual de instrucciones Pagina 54

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 61
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 53
Page 52 AD-4406 Weighing Indicator
14. RS-422/RS-485, Relay Output (OP-03)
Replacing the RS-232C interface with this option, the RS-422/RS-485 interface can
connect up to 32 indicators and control them from a computer or a PLC.
The functions of the RS-422/RS-485 interface are common to RS-232C except the
signal system.
The relays output the result of comparison.
Solid-state-relay
Maximum voltage 50VDC
Maximum current 100 mA DC
Maximum resistance 8 Ω
Pin connections
Function Pin No. Signal name Direction Description
1 SDA Output Transmission A terminal
2 SDB Output Transmission B terminal
3 RDA Input Receive A terminal
4 RDB Input Receive B terminal
5 TRM -
Terminator resistance (100 Ω)
RS-422
RS-485
6 SG - Signal ground
7 HI Output Relay output HI
8 OK Output Relay output OK
9 LO Output Relay output LO
Relay
output
10 COM - Relay output common
Adaptable connector TM-BLA10 (an accessory)
An AC adapter (an accessory) must be connected to the option board when using the
RS-422 or RS-485 interface. If using relay output only, the AC adapter is not required.
Circuit
14. RS-422/485, Relay Output (OP-03)
HI OK LO COM
HI
OK
LO
Relay Output Terminals
RS-422 Output
RS-485 Output
RS-422 Input
RS-485 Input
78 9
SDA SDB RDA RDB TRM
5V
47kΩ
4.7kΩ
47kΩ
4.7kΩ
100Ω
RS-422/RS-485 Terminals
1 2 3 4 5 6
SG
10
DC
DC 7V - 10V
AC adapter Jack
(AC adapter is an accessory of OP-03)
Note: Please confirm that the AC adapter type is correct
for your local voltage and receptacle type.
+
-
Vista de pagina 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 60 61

Comentarios a estos manuales

Sin comentarios