5. SMART RANGE FUNCTION
The AD-4212C-301/3100 are equipped with two ranges. The precision range has a higher resolution.
The standard range has a normal resolution.
The range switches automatically, depending on the value displayed. Sending the R command allows
weighing in the precision range, regardless of the tare value.
The range can be fixed to the standard range by changing the minimum weighing value using the
SMP command.
1 Send the R command
The balance will start weighing, using the
precision range.
2 Place a container on the weighing pan.
The weight value exceeds the precision range
value and the balance will switch to the
standard range.
3 Send the R command
The balance will switch to the precision range.
4 Place a sample in the container.
The weight value is within the precision range
value and the balance will perform a weighing,
using the precision range.
Precision range/standard range
Model Range used Weighing range
Available minimum
weighing value
Precision range
*1
0 g
to 51 g 0.0001 g
AD-4212C-301
Standard range
51 g to 320 g 0.001 g
Precision range
*2
0 g
to 510 g 0.001 g
AD-4212C-3100
Standard range
510 g to 3200 g 0.01 g
The factory setting is the precision range.
*1 When the weight value exceeds 51 g while the minimum weighing value of 0.0001 g
(precision range) is used, the balance will switch to the minimum weighing value of 0.001
g (standard range). By sending the R command, the balance can weigh from that point up
to 51 g, using the minimum weighing value of 0.0001 g (precision range).
*2 When the weight value exceeds 510 g while the minimum weighing value of 0.001 g
(precision range) is used, the balance will switch to the minimum weighing value of 0.01 g
(standard range). By sending the R command, the balance can weigh from that point up to
510 g, using the minimum weighing value of 0.001 g (precision range).
Display example: AD-4212C-301
Standard range
Precision range
Precision range
Precision range
Container
Weighing