Page 16 AD-4407 Weighing Indicator
Weighing a Mass
Weighing a Mass Weighing a Mass
Weighing a Mass except the
except the except the
except the Maximum Capacity
Maximum CapacityMaximum Capacity
Maximum Capacity
Step12 Specify a mass value using 0 - 9 keys.
Step13 Place a mass equivalent to displayed value
on the weighing unit. Proceed to step 15.
Weighing Capacity Mass
Step14 Place a mass equivalent to the maximum
capacity on the weighing unit.
Step15 Press the ENTER key after the STABLE
mark turns on. Proceed to step 17.
ESC key .......... The key to proceed to step17
without changing the span
parameters.
Caution Do not press the ENTER key while the
STABLE mark is off (detecting motion).
Arrange the condition using the F00
filter function.
Digital Input
Step16 Pressing the +/- key, a stored input voltage
parameter of the span is displayed in the
unit of mV/V. Adjust the input voltage
using the 0 - 9 keys. (It is possible to store
a greater value than the capacity.)
ENTER key...... The key to store the span
parameters and proceed to
step 17.
ESC key .......... The key to proceed to step
17 without changing the span
parameters.
5.2.5. Exiting the Calibration Mode
Step17 Check the Calend display.
Use the following keys.
CAL key ..... To store the parameters and display
Caloff. Proceed to the next step.
Caloff could be displayed by pressing
the +/- key while pressing the ON/OFF key
instead of CAL key.
5.2. Calibration Procedure
Turn on STABLE mark
Span calibration
Maximum capacity
To step 14, step 16
To step 12
New mass value
Mass o
f
maximum capacity
To step 14
To step 16
Place the displa
y
ed
mass on the
weighing unit
step 13
ste
p
14
Stored input voltage
New input voltage
mV/V
mV/V
Turn display off
9
0
9
0
+/
+/+/
+/
-
--
-
NET
/
GROS
S
PRINT
ENTER
PRIN
T
ENTER
PRINT
ENTER
+/
+/+/
+/
-
--
-